10. AMBIO Nikon Widefield Ti2

10. AMBIO Nikon Widefield Ti2

Description

General

Systems is equipped for multiplexed 3D live-cell experiments:

  • Multi-color, multi-position, time series, large image scan & automation
  • Fully motorized high precision stage for accurate repositioning (XYZ-control)
  • Stage incubator (temperature, CO2, humidity & carbogen)
    • Perfusion
    • Focus drift correction (Nikon PFS)
    • Water immersion objectives for high-resolution on water-based samples
    • Imaging Modes

      • Widefield epifluorescence (DAPI, CFP, GFP, YFP, mCherry & Cy5)
      • Bright-field (transmitted light) for histological slides (color staining)
      • Phase microscopy automated for 10x & 20x objectives
        • Interference Reflection Microscopy (IRM)
        • Polarized light microscopy
        • FRET
        • LED Illumination (Lumencore, SpectraX) and max. intensities out of 40x objective

          • 395 nm (395/25): 40 mW
          • 440 nm (438/29): 69 mW
          • 470 nm (475/28): 37 mW
          • 508 nm (511/16): 8 mW
          • 555 nm (555/28): 108 mW
          • 640 nm (635/22): 26 mW
          • Emission Filters

            Filterλ/Bandwidth
            Quad F67-401NXL (DAPI, GFP, mCherry, Cy5) 435/33
            520/26
            595/38
            695/63
            CFP ff01-482/35-32 482/35
            YFP ET540/30, F47-543 540/30
            GFP ET519/26M, F47-519NXL 519/26
            mCherry BrightLight 595/31 595/31
            Cy5 ET697/60M, F47-697NXL 697/60

            Detectors

            • For fluorescent samples: sCMOS, PCO.edge camera (>80% QE, 2048x2048 pixels, 6.5 µm pixel size, 53 fps full frame, up to 500 fps in small area)
            • For stained samples (histology): Nikon DS-Ri2 color CMOS camera (4908x3264 pixels, 6 fps full frame, 45 fps in 1636x1088 pixels)
            • Objectives

              Mag.TypeImmer.NAWD [μm]Info
              4x PlanApo λ Air 0.2 20,000 OFN25, ∞/-
              10x PlanFluo Ph1 DL Air 0.3 15,200 OFN25, ∞/1.2
              20x PlanApo Ph2 DM Air 0.75 1,000 ∞/0.17
              40x Plan Apo λ DIC N2 Air 0.95 250-170 OFN25, ∞/0.11-0.23
              40x Apo LWD λS DIC N2 Water 1.15 590-610 ∞/0.15-0.19
              60x Plan Apo λ DIC N2 Oil 1.4 130 OFN25, ∞/0.17

              Special

              • Fast LED switching (triggered acquisition)
              • Fast emission filter wheel for clean multi-color imaging of DAPI, CFP, GFP, YFP, mCherry & Cy5
              • Automatic water immersion dispenser (WID) for water immersion objectives
              • NIS-Elements AR (Advanced Research) software including modules for: JOBS (automation), 3D Deconvolution & NIS.ai (artificial intelligence): denoise, clarify, enhance, segment & convert
              Comments
              New users - access upon training request here.
              CONTACTS
              ambio@charite.de
              AFFILIATED ORGANIZATIONS
              Charité Universitätsmedizin Berlin